본문으로 바로가기
설치 보고서
DVIA-ULF Series
06-25-2020

KIST ZIESS-EXAMINER.A1 UD350 P7 Installation Report (June 2020)

DVIA-U Series
Installation Report
KIST
DVIA-UD350
Zeiss
ZIESS-EXAMINER.A1
2020
P7
Seoul

한국과학기술연구원 (UD350 P7)

현장 확인 항목

Target equipmentZIESS-EXAMINER.A1
Place of installationL7동 2층
RemarkGood Performance
END USER한국과학기술연구원

출장·작성 정보

항목내용
OperatorChoi, Hyoung Mun
Date of setup20.06.24
Date of reporting20.06.25
Date of business trip17.12.27~17.12.29

개요

After Active mounting, measure the vibration environment to check that there is no problem in using the equipment.

Measurement date: 2020.06.24 (WED),

Operator: Daeil System, Manager Choi, Hyoung Mun

Place of measurement : L7동 2층

Equipment Model: ZIESS-EXAMINER.A1

측정 조건 및 결과

DAQ System : B&K PULSE 14 & LAN-XI Type 3056

Accelerometer : PCB PIEZOTRONICS 393B05

Frequency Range : 0.7 ~ 200Hz

측정 결과 요약

Test ConditionVC-ClassZXY
FloorBFE
Above ActiveFGG

Good omnidirectional performance.

UD350 internal program performance measurement result

Z축 (내부 프로그램)

X축 (내부 프로그램)

Y축 (내부 프로그램)

Octave Band Result (BASE: Floor, TOP: Above Active)

Z축 (옥타브 밴드)

X축 (옥타브 밴드)

Y축 (옥타브 밴드)