Thermo Fisher Scientific Helios G5 DVIA-M1000 (230821R3-2) Installation Report

DVIA-M Series
Installation Report
Thermo Fisher Scientific
Helios G5
DVIA-M1000
230821R3-2

보고 조직

广电计量

M1000_ 230821R3-2 Set up

양식 정보

Target equipmentThermo Fisher Helios G5
Place of installationWuxi
RemarkGood Performance
END USER广电计量
Operator: Lee, Chaewon Date of setup: 24.03.20 Date of reporting: 24.03.21Date of business trip : 17.12.27~17.12.29

개요

The DVIA-M1000 active vibration isolation platform is appropriately installed and functioning as intended. The center of gravity of the equipment above the product is not aligned, so a resonance peak occurs around 34Hz. Measurement date: 2024.03.20 (Wednesday) Operator: DAEIL SYSTEMS Chaewon Lee Place of measurement: 无锡 Equipment Model: Thermo Fisher Helios G5

M1000 내부 프로그램 성능 측정 결과

수직

좌우

전후