설치 보고서
DVIA-M Series
06-22-2022
Thermo Scientific Axia DVIA-M1000 Installation Report (Jiaxing, 220106R9, 2022-06-22)
DVIA-M Series
Installation Report
DVIA-M1000
Thermo Scientific
Axia
Jiaxing
220106R9
China
Internal program
End user (표지)
浙江长盛滑动轴承股份有限公司 / ZHEJIANG CHANGSHENG SLIDING BEARING CO., LTD
플랫폼 제목
(M1000_ 220106R9 Set up)
셋업·타이틀 화면
대상 장비
thermoscientific Axia
설치 장소
Jiaxing, China
비고
Good Performance
End user (표)
Zhejiang Changsheng Sliding Bearing Co. Ltd.(浙江长盛滑动轴承股份有限公司)
Operator
Yu, Ho Sang
설치일
22.06.22
보고서 작성일
22.06.22
출장 일정
17.12.27~17.12.29
개요
The DVIA-M1000 active vibration isolation platform is appropriately installed and functioning as intended.
측정일
2022.06.22 (Wednesday),
측정 Operator
DAEIL SYSTEMS, Hosang Yu
측정 장소
jiaxing, China
장비 모델
thermoscientific Axia
측정 조건 및 결과
측정 결과 요약
Test Condition
VC-Class
| Z | X | Y | |
|---|---|---|---|
| Floor | B | B | B |
| Above Active | F | E | F |
Good omnidirectional performance.
MB1000 internal program performance measurement result
Z axis
X axis
Y axis
Octave Band Result
Z axis
X axis
Y axis
Reference
일반 진동 기준
참고:
- VC-A/B는 8-80 Hz, VC-C~VC-G는 1-80 Hz 범위의 1/3 옥타브 밴드에서 측정.
- 세부 크기는 마이크로일렉트로닉스 제조에서의 선폭 또는 의료 연구에서의 입자 크기를 의미.
이 케이스 스터디 공유하기
케이스 스터디 정보
카테고리
설치 보고서
시리즈DVIA-M Series
작성일06-22-2022
태그
DVIA-M Series
Installation Report
DVIA-M1000
Thermo Scientific
Axia
Jiaxing
220106R9
China
Internal program
관련 케이스 스터디

DVIA-M Series
Jiangsu Modern Road & Bridge Axia ChemiSEM HiVac DVIA-M1000 (220718R3) Installation Report — Nanjing
09-19-2022자세히

DVIA-M Series
Jiaxing Xindong Technology ZEISS Sigma 360 FE-SEM DVIA-M1000 (250318R5) Installation Report
06-09-2025자세히

DVIA-M Series
ZEISS EVO 10 DVIA-M1000 (191211R3) installation worksheet — 2025.05.14
05-14-2025자세히