설치 보고서
DVIA-M Series
08-05-2022
Ningbo Institute of materials ZEISS GEMINI300 DVIA-M1000 (210714R2-2) Installation Report — 2022.08.05
DVIA-M Series
Installation Report
Ningbo Institute of materials
ZEISS
GEMINI300
SEM
DVIA-M1000
210714R2-2
목차
Ningbo Institute of materials(宁波材料所)
(M1000_ 210714R2-2 Set up)
| Target equipment | ZEISS GEMINI300 |
|---|---|
| Place of installation | BINZHOU |
| Remark | Good Performance |
| END USER | Ningbo Institute of materials(宁波材料所) |
| Operator: Lee, Young ha Date of setup: 22.08.04 Date of reporting: 22.08.05 | Date of business trip : 17.12.27~17.12.29 |
개요
The DVIA-M1000 active vibration isolation platform is appropriately installed and functioning as intended.
측정 일자
Measurement date: 2022.08.04 (Thursday),
담당자
Operator: DAEIL SYSTEMS, Youngha Lee
측정 장소
Place of measurement: BINZHOU
장비 모델
Equipment Model: ZEISS GEMINI300
측정 조건 및 결과
측정 결과 요약
Test Condition
VC-Class
| Z | X | Y | |
|---|---|---|---|
| Floor | B | E | E |
| Above Active | G | F | F |
Good omnidirectional performance.
MB1000 내부 프로그램 성능 측정 결과
MB1000 내부 프로그램 — Z축
MB1000 내부 프로그램 — X축
MB1000 내부 프로그램 — Y축
옥타브 밴드 결과
옥타브 밴드 결과 — Z축
옥타브 밴드 결과 — X축
옥타브 밴드 결과 — Y축
이 케이스 스터디 공유하기
케이스 스터디 정보
카테고리
설치 보고서
시리즈DVIA-M Series
작성일08-05-2022
태그
DVIA-M Series
Installation Report
Ningbo Institute of materials
ZEISS
GEMINI300
SEM
DVIA-M1000
210714R2-2
관련 케이스 스터디

DVIA-M Series
Ningbo Institute of materials ZEISS Gemini 300 DVIA-M1000 (210714R2-2) Internal Program Performance — Binzhou — 2022.07.27
07-27-2022자세히

DVIA-M Series
Chinese Academy of Sciences — Institute of Metal Research ZEISS GeminiSEM360 DVIA-M1000 (230328R2) Installation Report
05-15-2026자세히

DVIA-M Series
YIMO Zeiss Technology (Suzhou) ZEISS EVO 10 SEM DVIA-M1000 (240304R8) Installation Report
11-11-2025자세히