3Trees 코팅 Thermo Scientific Apreo 2C DVIA-MB1000(210705R2) 내부 프로그램 성능 — 2021-11-29
DVIA-M Series
Installation Report
YIMO
DVIA-MB1000
Thermo Scientific
Apreo 2C
210705R2
3Trees Coatings
목차
(MB1000 P253 210705R2 Set up)
| Target equipment | Thermo Apreo 2C |
|---|---|
| Place of installation | N/A |
| Remark | Good Performance |
| END USER | 三棵树涂料股份有限公司 |
| Operator: Choi, HyoungMun | Date of setup : 21.11.25 |
장비 사진
개요
After Active mounting, measure the vibration environment to check that there is no problem in using the equipment.
Measurement date: 21.11.25 (thu)
Operator: Daeil System, Research engineer Choi, HyoungMun
Place of measurement : N/A
Equipment Model: Thermo Apreo 2C
측정 결과 요약
| Test Condition | 2 Hz | 5 Hz | 10 Hz |
|---|---|---|---|
| Z | -30 dB | -27 dB | -25 dB |
| X | -40 dB | -35 dB | -32 dB |
| Y | -40 dB | -32 dB | -20 dB |
Good omnidirectional performance.
MB1000 internal program performance measurement result
Z축
X축
Y축