대상 장비 및 설치 장소

Target equipmentTescan solaris gmh
Place of installationN/A
RemarkGood Performance
END USERHuawei

보고 일정

Operator: Kim, Doohyun

Date of setup : 21.09.24

Date of reporting : 21.09.29

Date of business trip : 17.12.27~17.12.29

개요

After Active mounting, measure the vibration environment to check that there is no problem in using the equipment.

Measurement date: 21.09.24 (FRI)

Operator: Daeil System, Research engineer Kim, Doohyun

Place of measurement : N/A

Equipment Model: Tescan solaris gmh

측정 조건 및 결과

측정 결과 요약

시험 조건

전달률

Test ConditionTransmissibilityTransmissibilityTransmissibility
2 Hz5 Hz10 Hz
Z-23 dB-40 dB-40 dB
X-29 dB-35 dB-31 dB
Y-37 dB-33 dB-31 dB

Good omnidirectional performance.

MB1000 내부 프로그램 성능 측정 결과

Z axis

X axis

Y axis