설치 보고서
DVIA-M Series
01-22-2021
BOE ZEISS Crossbeam 550 DVIA-MB1000 (MB1000 P181, YIMO 200720R5-1) Internal Program — 2021.01.22
DVIA-M Series
Installation Report
BOE
ZEISS
Crossbeam 550
DVIA-MB1000
MB1000 P181
YIMO
200720R5-1
목차
(MB1000 P181 YIMO 200720R5-1)
| Target equipment | ZEISS-Crossbeam550 |
|---|---|
| Place of installation | N/A |
| Remark | Good Performance |
| END USER | BOE |
| Operator: Choi, HyoungMun | Date of business trip : 17.12.27~17.12.29 |
| Date of setup : 21.01.22 | Date of reporting : 21.01.22 |
개요
After Active mounting, measure the vibration environment to check that there is no problem in using the equipment.
Measurement date: 21.01.22 (FRI)
Operator: Daeil System, Research engineer Choi, HyoungMun
Place of measurement : N/A
Equipment Model: ZEISS-Crossbeam550
측정 조건 및 결과
측정 결과 요약
| Test Condition | Transmissibility | ||
|---|---|---|---|
| 2 Hz | 5 Hz | 10 Hz | |
| Z | -23 dB | -28 dB | -33 dB |
| X | -23 dB | -32 dB | -34 dB |
| Y | -25 dB | -32 dB | -40 dB |
Good omnidirectional performance.
MB1000 internal program performance measurement result
Z축
X축
Y축
이 케이스 스터디 공유하기
케이스 스터디 정보
카테고리
설치 보고서
시리즈DVIA-M Series
작성일01-22-2021
태그
DVIA-M Series
Installation Report
BOE
ZEISS
Crossbeam 550
DVIA-MB1000
MB1000 P181
YIMO
200720R5-1
관련 케이스 스터디

DVIA-M Series
YIMO ZEISS CROSSBEAM350 DVIA-MB1000 (210705R1 · MB1000 P244) Internal Program Performance — 2021.11.05
11-05-2021자세히

DVIA-M Series
ZEISS Crossbeam 550 DVIA-MB1000 P228 (210510R5) Internal Program Performance — YIMO — 2021.10.28
10-28-2021자세히

DVIA-M Series
Shagang Group ZEISS Crossbeam 550 DVIA-MB1000 (201028R1 · P202, YIMO remote tuning) — 2021.04.12
04-12-2021자세히