설치 보고서
DVIA-M Series
04-21-2021
Novartis Thermo Fisher Scientific Glacios DVIA-MB3000 (201012R5) Installation Report — 2021.04.21
DVIA-M Series
Installation Report
Novartis
Thermo Fisher Scientific
Glacios
DVIA-MB3000
201012R5
VEC
Contents
장비 사진
대상 장비
TFS Glacios
설치 장소
Novartis, MA
비고
Good Performance
END USER
Novartis
Operator:
Doohyun Kim
Date of setup :
21.04.21
Date of reporting :
21.04.22
Date of business trip :
17.12.27~17.12.29
개요
능동 마운팅 후 진동 환경을 측정하여 장비 사용에 문제가 없음을 확인합니다.
Measurement date:
21.04.21 (Wed)
Operator:
Daeil System, Research engineer Doohyun Kim
Place of measurement :
Novartis, MA
Equipment Model:
TFS Glacios
측정 조건 및 결과
측정 결과 요약
| Test Condition | 2 Hz | 5 Hz | 10 Hz |
|---|---|---|---|
| Transmissibility — Z | -16 dB | -23 dB | -30 dB |
| Transmissibility — X | -23 dB | -15 dB | -25 dB |
| Transmissibility — Y | -30 dB | -25 dB | -26 dB |
| Test Condition | Floor | Top Plate | |
| VC Class — Z | D | F | |
| VC Class — X | G | G | |
| VC Class — Y | G | G |
Good omnidirectional performance.
MB1000 내장 프로그램 성능 측정 결과
Z axis
X axis
Y axis
VEC 측정 데이터
Z Axis
X Axis
Y Axis
이 케이스 스터디 공유하기
케이스 스터디 정보
카테고리
설치 보고서
시리즈DVIA-M Series
작성일04-21-2021
태그
DVIA-M Series
Installation Report
Novartis
Thermo Fisher Scientific
Glacios
DVIA-MB3000
201012R5
VEC
관련 케이스 스터디

DVIA-M Series
Seoul National University Thermo Fisher Scientific Glacios DVIA-MB1000 (210121R5) Installation Report
03-31-2026자세히

DVIA-M Series
Changping P3 Laboratory Thermo Fisher Scientific Cryo-TEM Krios G4 DVIA-MB3000 (250708R5-1) Installation Report — November 5, 2025
11-05-2025자세히

DVIA-M Series
Hainan University Thermo Fisher Spectra Ultra DVIA-MB3000 (231004R3-3) Installation Report
04-18-2024자세히