본문으로 바로가기
설치 보고서
DVIA-M Series
08-10-2020

Beijing University of Aeronautics and Astronautics FEI Scios2 DVIA-MB1000 (190925R1 · P143) Remote measurement — Internal program plots — 2020.08.10

DVIA-M Series
Installation Report
DVIA-MB1000
190925R1
FEI
Scios2
Beijing
YIMO
remote
P143

(MB1000_190925R1(P143) Set up)

Target equipmentFEI Scios2
Place of installationBeijing University of Aeronautics and Astronautics, 4th floor
RemarkGood Performance
END USERBeijing University
Operator: Kim, DoohyunDate of business trip : 17.12.27~17.12.29
Date of setup : 20.08.05.Date of reporting : 20.08.10

개요

After Active mounting, measure the vibration environment to check that there is no problem in using the equipment.

Measurement date: 20.08.05 (WED),

Operator: Daeil System, Manager Doohyun, Kim

Place of measurement : Beijing University of Aeronautics and Astronautics 4th floor

Equipment Model: FEI Scios2

측정 조건 및 결과

측정 결과 요약

Test ConditionVC-Class
ZXY
Floor
Above Active

Good omnidirectional performance.

MB1000 internal program performance measurement result

Z축

X축

Y축

이 케이스 스터디 공유하기

케이스 스터디 정보

카테고리
설치 보고서
시리즈DVIA-M Series
작성일08-10-2020
태그
DVIA-M Series
Installation Report
DVIA-MB1000
190925R1
FEI
Scios2
Beijing
YIMO
remote
P143