남중국이공대학교 Hitachi Regulus 8100 DVIA-MB1000(190117R3) 설치 보고서 — 2019.07.30
DVIA-M Series
Installation Report
DVIA-MB1000
MB1000
South China University of Technology
Hitachi Regulus
190117R3
Guangzhou
Contents
설치 사진
Target equipment
Hitachi Regulus 8100
Place of installation
South China University of Technology F1
Remark
Good Performance
END USER
South China University of Technology
Operator: Choi, Sung Won
Date of setup : 19.07.24
Date of reporting : 19.07.30
Date of business trip : 17.12.27~17.12.29
개요
After Active mounting, measure the vibration environment to check that there is no problem in using the equipment.
Measurement date: 2019.07.24 (wednesday),
Operator: Daeil System, Manager Sung Won Choi
Place of measurement : 1st Floor, South China University of Technology
Equipment Model: Hitachi Regulus 8100
측정 조건 및 결과
측정 결과 요약
Test Condition
VC-Class
| Test Condition | VC-Class | ||
|---|---|---|---|
| Z | X | Y | |
| Floor | D | D | D |
| Above Active | F | E | G |
Good omnidirectional performance.
MB1000 내부 프로그램 성능 측정 결과
Z 축 (MB1000 내부 프로그램)
X 축 (MB1000 내부 프로그램)
Y 축 (MB1000 내부 프로그램)
옥타브 밴드 결과 (BASE: Floor, TOP: Above Active)
Z 축 (옥타브 밴드)
X 축 (옥타브 밴드)
Y 축 (옥타브 밴드)