Fudan University OXFORD AS400 NMR DVIA-MB1000 (MB1000 program P75) Performance Report — 2018.04.10
DVIA-M Series
Installation Report
Fudan University
Oxford Instruments
NMR
DVIA-MB1000
MB1000
P75
목차
Pudan Univ (MC1000_P75 Set up)
Target equipment
| Target equipment | OXFORD AS400 NMR |
|---|---|
| Place of installation | Pudan UNIV, Basement 1FLOOR |
| Uniqueness | Good Performance |
| END USER | Pudan Univ |
| Date of business trip : 18.03.30 | Business traveler: Choi, HyoungMun |
| Date Created : 18.04.10 |
개요
After Active mounting, measure the vibration environment to check that there is no problem in using the equipment.
측정 일자
Measurement date: 2018.03.30 (Friday), Rater: Daeil System, Manager Choi , Hyoung Moon
측정 장소
Place of measurement : Pudan UNIV, Basement 1FLOOR
Equipment Model
Equipment Model: OXFORD AS400 NMR
측정 조건 및 결과
N/A
측정 결과 요약
| Test Condition | VC-Class Z | VC-Class X | VC-Class Y |
|---|---|---|---|
| Floor | E | E | E |
| Above Active | G | G | G |
Good omnidirectional performance and attached VC-Class graph.
MB1000 internal program performance measurement result
Z axis - internal program
X axis - internal program
Y axis - internal program
Octave Band Result (BASE: Floor, TOP: Above Active)
Z axis - octave band result
X axis - octave band result
Y axis - octave band result
이 케이스 스터디 공유하기
케이스 스터디 정보
카테고리
설치 보고서
시리즈DVIA-M Series
작성일04-10-2018
태그
DVIA-M Series
Installation Report
Fudan University
Oxford Instruments
NMR
DVIA-MB1000
MB1000
P75
관련 케이스 스터디

DVIA-M Series
Fudan University Bruker D8 VENTURE DVIA-MB1000 (MB1000 program P60) Performance Report — 2018.04.10
04-10-2018자세히

DVIA-M Series
YIMO Pudan University site — MC3000 / MC1000 / MB1000 vibration setup report (2018-01-30)
01-30-2018자세히

DVIA-M Series
Shanghai Jiaotong University Thermo Fisher Apreo 2S DVIA-MB1000 (210701R6) Installation Report — 2024.07.06
07-06-2024자세히