설치 보고서
DVIA-M Series
04-10-2018
FEI Nova NanoSEM450 DVIA-MB1000 (MB1000 P61) Performance Report — 2018.04.10
DVIA-M Series
Installation Report
DVIA-MB1000
MB1000
P61
FEI
Nova NanoSEM450
Pudan UNIV
목차
(Pudan Univ (MB1000_P61 Set up))
Target equipment
| Target equipment | FEI Nova NanoSEM450 |
|---|---|
| Place of installation | Pudan UNIV, Basement 1FLOOR |
| Uniqueness | Good Performance |
| END USER | DONGJI UNIV |
| Date of business trip : 18.03.30 | |
| Business traveler: Choi, HyoungMun | |
| Date Created : 18.04.10 |
개요
After Active mounting, measure the vibration environment to check that there is no problem in using the equipment.
측정 일자
Measurement date: 2018.03.30 (Friday), Rater: Daeil System, Manager Choi , Hyoung Moon
측정 장소
Place of measurement : Pudan UNIV, Basement 1FLOOR
Equipment Model
Equipment Model: FEI Nova NanoSEM450
측정 조건 및 결과
N/A
측정 결과 요약
| Test Condition | VC-Class Z | VC-Class X | VC-Class Y |
|---|---|---|---|
| Floor | E | E | E |
| Above Active | E | G | F |
Good omnidirectional performance and attached VC-Class graph.
MB1000 internal program performance measurement result
Z axis - internal program
X axis - internal program
Y axis - internal program
Octave Band Result (BASE: Floor, TOP: Above Active)
Z axis - octave band result
X axis - octave band result
Y axis - octave band result
Reference
Generic Vibration Criteria
참고:
1. VC-A/B는 8–80 Hz, VC-C~G는 1–80 Hz의 1/3 옥타브 밴드에서 측정된 값입니다.
2. Detail size는 반도체 제조의 폭 또는 의학 연구의 입자 크기를 가리킵니다.
이 케이스 스터디 공유하기
케이스 스터디 정보
카테고리
설치 보고서
시리즈DVIA-M Series
작성일04-10-2018
태그
DVIA-M Series
Installation Report
DVIA-MB1000
MB1000
P61
FEI
Nova NanoSEM450
Pudan UNIV
관련 케이스 스터디

DVIA-M Series
FEI NOVA450 DVIA-MB1000 (MB1000 P8) installation worksheet — Seoul Hanyang University — 2020-07-14
07-14-2020자세히

DVIA-M Series
FEI SEM DVIA-MB1000 Installation Report — MB1000 P142, SN 190916R4 (2020.06.04)
06-04-2020자세히

DVIA-M Series
Xiamen University FEI QUANTA FEG 650 DVIA-MB1000 (MB1000 P62) Internal program worksheet — 2018-09-17
09-17-2018자세히