설치 보고서
DVIA-M Series
12-27-2017
Kolbenschmidt Shanghai Piston ZEISS evo18 DVIA-MB1000 (P56) Internal program worksheet — 2017.12.27
DVIA-M Series
Installation Report
YIMO
DVIA-MB1000
P56
Zeiss
evo18
Kolbenschmidt
Shanghai
목차
Kolbenschmidt Shanghai Piston
(MB1000_P56 Set up)
Target equipment
ZEISS-evo18
Place of installation
1th Floor, Kolbenschmidt Shanghai Piston
Uniqueness
Good performance
END USER
Kolbenschmidt Shanghai Piston
Date of business trip
Date of business trip : 17.12.27~17.12.29
Business traveler
Business traveler: Choi, SungWon
Date Created
Date Created : 18.01.02
개요
After Active mounting, measure the vibration environment to check that there is no problem in using the equipment.
측정 일자
Measurement date: 2017.12.27 (Monday), Rater: Daeil System, Manager Sung Won Choi
측정 장소
Place of measurement : 1th Floor, Kolbenschmidt shanghai piston
Equipment Model
Equipment Model: ZEISS-evo18
측정 조건 및 결과
측정 결과 요약
| 테스트 조건 | Z | X | Y |
|---|---|---|---|
| Floor | D | E | E |
| Above Active | E | F | E |
Good omnidirectional performance and attached VC-Class graph.
VC-class notation except for 70Hz noise
MB1000 내부 프로그램 성능 측정 결과
Z axis
X axis
Y axis
Octave Band Result (BASE: Floor, TOP: Above Active)
옥타브 밴드 — Z axis
옥타브 밴드 — X axis
옥타브 밴드 — Y axis
이 케이스 스터디 공유하기
케이스 스터디 정보
카테고리
설치 보고서
시리즈DVIA-M Series
작성일12-27-2017
태그
DVIA-M Series
Installation Report
YIMO
DVIA-MB1000
P56
Zeiss
evo18
Kolbenschmidt
Shanghai
관련 케이스 스터디

DVIA-M Series
YIMO Zeiss FE-SEM Sigma360 DVIA-M1000 (250102R2-4) Installation Report
04-30-2025자세히

DVIA-M Series
YIMO Zeiss FE-SEM Sigma360 DVIA-M1000 (250102R2-4) Installation Report
04-18-2025자세히

DVIA-M Series
Huawei (Hitachi) Hitachi SU8200 DVIA-MB1000 (210601R1 YIMO / P251) Internal Program Performance — Shanghai — 2021.12.07
12-07-2021자세히