본문으로 바로가기
설치 보고서
DVIA-M Series
05-27-2016

DVIA-M type (MB1000) installation test report — 2016-05-27

DVIA-M Series
Installation Report
DVIA-M1000
MB1000
Transmissibility
B&K PULSE
Daeil Systems
2016

DVIA-M type (MB1000) Installation Test Report

Installation & Mesurement Date : 2016.05.25.

Reporting Date : 2016.05.27

DAEIL SYSTEMS

LIM JONG UK

※ Installation Results Summary

테스트 방향VC-Class (Base — 바닥)VC-Class (Top — Active 상태 상부)
ACTIVEZCF
XDE
YCE

Good performance in all directions.

Vertical axis Z is C Class, the horizontal axis X is D Class, the horizontal axis Y is C Class. It was measured vibration environment in floor.

Vertical axis Z is F Class, the horizontal axis X is E Class, the horizontal axis Y is E Class. It was measured vibration environment in top of Active state.

Used data of internal program to making performance graph, show “Result 1”.

conversion unit in 1/3 octave band, Show “Result 2“.

we check image of equipment, Show “Result 3“.

Result 1 <Performance graph used data of internal program – Transmissibility >

Result 1: Z axis

Result 1: X axis

Result 1: Y axis

Result 2 <Performance graph used data of PULSE – 1/3 Octave band>

Result 2: Z axis

Result 2: X axis

Result 2: Y axis

Result 3 <Image Check>

Floor

Passive

Active