설치 보고서
DVIA-M Series
06-05-2014

DVIA-M1000 P3 factory performance test (2014)

DVIA-M Series
Installation Report
DVIA-M1000
Performance Test
Transmissibility
Daeil Systems
2014

Product Performance Test Report

Model : DVIA-M1000 P3 (서울대 설치 예정)

Purpose : Performance Test

Measurement Engineer : Sungwon Choi

Time and Place : Factory in DaeilSystem, 2014-6-4

Measurement : Vertical, Horizontal direction Transmissibility

Supporting figure after test scope

Test equipment

Data Acquisition system : VAAV-1000E monitoring program/3050-B-040

Accelerometer : GS-11D in DVIA-M units/23435,35151(PCB)

Frequency Range : 0.5~100Hz

Test Conditions

Load Weight :약 700kg(프레임 상판 자체중량 포함)

Isolation system : 1set

Inner Air Pressure : N/A

Measurement methods

Sensor Position : Center

Moving Direction : N/A

Supporting figure after measurement methods

Test Result (Transmissibility)

Table 1. Summary

Test DirectionResonance Frequency [Hz]Roll-off Rate [dB]
Z CenterN/A-28.3
X CenterN/A-36.3
Y CenterN/A-34.0

Roll-off Rate is evaluated at 10 Hz.

VAAV built-in program results — FFF Z

VAAV built-in program results — FFF X

VAAV built-in program results — FFF Y

Pulse measurement and cardboard test environment

Passive X traces

Issues and improvements

전 방향 1~20Hz 성능은 양호함

Y축은 고주파 성능이 좋지 않은 것으로 나타나는데, 튜닝에 따라 충분히 개선 가능한 상황이며, 면밀히 검토하지 않고 튜닝을 마친 것이 문제점.

더미 아래에 골판지가 깔려 있지 않았을 때는 패시브와 오픈루프 특성 상 고주파 영역에 이상 피크가 없었음.

골판지를 더미 아래에 깔고 나서 측정한 결과 X방향에서 85Hz 피크가 발생하여, M1000 P2에서 있었던 수평축 이상 피크의 원인이 골판지였음을 확인할 수 있었음. Y, Z 방향에서는 거의 영향을 발견할 수 없어, 골판지가 방향성 공진을 일으킨다는 다나카 부장의 의견이 타당하다고 생각됨.

이 케이스 스터디 공유하기

케이스 스터디 정보

카테고리
설치 보고서
시리즈DVIA-M Series
작성일06-05-2014
태그
DVIA-M Series
Installation Report
DVIA-M1000
Performance Test
Transmissibility
Daeil Systems