설치 보고서
DVIA-M Series
02-21-2023
Xi'an Jiaotong University ZEISS Crossbeam 550 FIB-SEM DVIA-M1000 (220614R7) Installation Report
DVIA-M Series
Installation Report
Xi'an Jiaotong University
ZEISS
Crossbeam 550
FIB-SEM
DVIA-M1000
220614R7
Contents
개요
XI'AN JIAOTONG UNIVERSITY
(M1000_ 220614R7 Set up)
| 대상 장비 | Crossbeam 550 |
|---|---|
| 설치 장소 | XI'AN |
| 비고 | Good Performance |
| END USER | XI'AN JIAOTONG UNIVERSITY |
Operator: Lee, Young Ha
Date of setup: 23.02.13
Date of reporting: 23.02.21
Date of business trip : 17.12.27~17.12.29
Overview
The DVIA-M1000 active vibration isolation platform is appropriately installed and functioning as intended.
Measurement date: 2023.02.13 (Monday), Operator: DAEIL SYSTEMS, Youngha Lee
Place of measurement: XI'AN
Equipment Model: Crossbeam 550
Measurement conditions and results
Summary of measurement results
Test Condition VC-Class Z X Y
Floor B E E
Above Active E D E
Good omnidirectional performance.
5.
장비
M1000 내부 프로그램 성능 측정 결과
수직 전달률 및 코히어런스
좌우 전달률 및 코히어런스
전후 전달률 및 코히어런스
옥타브 밴드 결과
수직 옥타브 밴드 결과
좌우 옥타브 밴드 결과
전후 옥타브 밴드 결과
Reference
일반 진동 기준
참고:
- VC-A/B는 8-80 Hz, VC-C~VC-G는 1-80 Hz 범위의 1/3 옥타브 밴드에서 측정.
- 세부 크기는 마이크로일렉트로닉스 제조에서의 선폭 또는 의료 연구에서의 입자 크기를 의미.
이 케이스 스터디 공유하기
케이스 스터디 정보
카테고리
설치 보고서
시리즈DVIA-M Series
작성일02-21-2023
태그
DVIA-M Series
Installation Report
Xi'an Jiaotong University
ZEISS
Crossbeam 550
FIB-SEM
DVIA-M1000
220614R7
관련 케이스 스터디

DVIA-M Series
YIMO BYD Semiconductor ZEISS Crossbeam 550 FIB-SEM DVIA-M1000 (240314R6) Installation Report
05-11-2026자세히

DVIA-M Series
YIMO ZEISS FIB-SEM Crossbeam 550 DVIA-M1000 (240417R5) Installation Report
05-11-2026자세히

DVIA-M Series
YIMO Tianjin Institute of Hematology ZEISS Crossbeam 550 FIB-SEM DVIA-M1000 (250530R5-2) Installation Report
08-25-2025자세히