설치 보고서
DVIA-ULF Series
05-26-2026
YIMO DVIA-UD350 (210929R1, P26) OLIMPUS remote tuning worksheet — UD350 internal Z / X / Y program plots
DVIA-U Series
Installation Report
YIMO
OLIMPUS
DVIA-UD350
210929R1
P26
remote tuning
목차
(UD350_210929R1 P26 Set up)
현장 확인 항목
| Target equipment | OLIMPUS |
|---|---|
| Place of installation | N/A |
| Remark | Good Performance |
| END USER | N/A |
운영자 및 일정
| Operator: CHOI, HYOUNGMUN<br>Date of setup : 21.12.15<br>Date of reporting : 21.12.17 | Date of business trip : 17.12.27~17.12.29 |
|---|
개요
After Active mounting, measure the vibration environment to check that there is no problem in using the equipment.
Measurement date: 21.12.15(WEN)
Operator: Daeil System, Research engineer YU, HoSang
Place of measurement : N/A
Equipment Model: OLIMPUS
측정 조건 및 결과
측정 결과 요약
Transmissibility
| Test Condition | 2 Hz | 5 Hz | 10 Hz |
|---|---|---|---|
| Z | -16 dB | -25 dB | -29 dB |
| X | -5 dB | -22 dB | -29 dB |
| Y | -11 dB | -22 dB | -29 dB |
Good omnidirectional performance.
UD350 내부 프로그램 성능 측정 결과
Z축
X축
Y축
참고자료
Generic Vibration Criteria
참고:
- VC-A/B는 8–80 Hz의 1/3 옥타브 밴드에서, VC-C~VC-G는 1–80 Hz에서 측정합니다.
- 세부 크기는 마이크로일렉트로닉스 제조의 선폭 또는 의료 연구에서의 입자 크기 등을 의미합니다.
이 케이스 스터디 공유하기
케이스 스터디 정보
카테고리
설치 보고서
시리즈DVIA-ULF Series
작성일05-26-2026
태그
DVIA-U Series
Installation Report
YIMO
OLIMPUS
DVIA-UD350
210929R1
P26
remote tuning
관련 케이스 스터디
DVIA-ULF Series
Lattice Semiconductor Phils ZEISS EVO 18 DVIA-U1000 (200203R4 · P15) Remote tuning — Philippines — 2020.10.08
10-08-2020자세히

DVIA-ULF Series
Liupan Mountain Laboratory ZEISS Sigma 360 FE-SEM DVIA-U700 (241115R3) Installation Report — Yinchuan
07-14-2025자세히

DVIA-ULF Series
Korea Institute of Science and Technology (L7-222) Olympus BX51 DVIA-UD350 (220127R1-2) Installation Report
12-31-2024자세히