(M1000_210701R6 Re-tuning)
대상 장비 및 설치 장소
| Target equipment | THERMO APREO 2S |
|---|---|
| Place of installation | GUANGZHOU, 1F |
| Remark | Good Performance |
| END USER | Opton |
보고서 세부 정보
| Operator: Yu, Ho Sang<br>Date of setup : 22.05.18<br>Date of reporting : 22.05.19 | Date of business trip : 17.12.27~17.12.29 |
|---|---|
| Overview The DVIA-M1000 active vibration isolation platform is appropriately installed and functioning as intended. Measurement date: 2022.05.18 (Wednesday), Operator: Daeil System, HoSang Yu Place of measurement: GUANGZHOU, 1F Equipment Model: THERMO APREO 2S |
측정 조건 및 결과
측정 결과 요약
| Test Condition | Transmissibility | ||
|---|---|---|---|
| 2 Hz | 5 Hz | 10 Hz | |
| Z | -24.13 dB | -21.34 dB | -26.32 dB |
| X | -21.09 dB | -21.30 dB | -25.45 dB |
| Y | -21.58 dB | -20.90 dB | -25.57 dB |
Good omnidirectional performance.
MB1000 내부 프로그램 성능 측정 결과
Z축
X축
Y축