본문으로 바로가기
설치 보고서
DVIA-M Series
08-24-2021

YIMO Thermo Fisher Scientific APREO 2S DVIA-MB1000 (210701R6-1 P241) Installation Report — 2021.08.24

DVIA-M Series
Installation Report
YIMO
DVIA-MB1000
210701R6-1
P241
Thermo Fisher Scientific
APREO 2S

MB1000 P241(210701r6-1)

대상 장비

THERMO APREO 2S

설치 장소

N/A

비고

Good Performance

최종 사용자

N/A

Operator

Kim, Doohyun

Date of setup

21.08.24

Date of reporting

21.08.24

Date of business trip

17.12.27~17.12.29

개요

After Active mounting, measure the vibration environment to check that there is no problem in using the equipment.

Measurement date: 21.08.24 (TUE)

Operator: Daeil System, Research engineer Kim, Doohyun

Place of measurement: N/A

Equipment Model: THERMO APREO 2S

진동 제어 시스템 정보

모델: DVIA-MB1000 (P241)

시리얼 넘버: 210701r6-1

측정 조건 및 결과

측정 결과 요약

Transmissibility

2 Hz5 Hz10 Hz
Z-21 dB-30 dB-32 dB
X-32 dB-35 dB-31 dB
Y-29 dB-40 dB-25 dB

Good omnidirectional performance.

MB1000 internal program performance measurement result

Z axis

X axis

Y axis

이 케이스 스터디 공유하기