TPK 홀딩 Thermo Fisher Helios G5 DVIA-MB1000(201214R4 · P216) 원격 성능평가 — 내부 프로그램 플롯 — 2021.04.09
DVIA-M Series
Installation Report
DVIA-MB1000
201214R4
Thermo Fisher Scientific
Helios G5
Xiamen
YIMO
목차
(MB1000_201214R4 (P216) Set up)
| Target equipment | THERMO FISCHER Helios G5 |
|---|---|
| Place of installation | Xiamen, 4F, TPK Holding |
| Remark | Good Performance |
| END USER | TPK Holding Co., Ltd |
| Operator: Doohyun Kim | Date of setup : 21.04.08 |
| Date of reporting : 21.04.09 | Date of business trip : 17.12.27~17.12.29 |
개요
After Active mounting, measure the vibration environment to check that there is no problem in using the equipment.
Measurement date: 21.04.08 (Thu)
Operator: Daeil System, Research engineer Doohyun Kim
Place of measurement : Xiamen, 4F, TPK Holding
Equipment Model: THERMO FISCHER Helios G5
측정 조건 및 결과
측정 결과 요약
| 2 Hz | 5 Hz | 10 Hz | |
|---|---|---|---|
| Z | -28 dB | -25 dB | -32 dB |
| X | -27 dB | -30 dB | -31 dB |
| Y | -25 dB | -24 dB | -32 dB |
Good omnidirectional performance.
MB1000 internal program performance measurement result
Z축
X축
Y축