설치 보고서
DVIA-M Series
01-10-2022
YIMO UB1000 P24 (200828R4) ZEISS Sigma 300 remote tuning measurement report
DVIA-M Series
Installation Report
YIMO
Tier-1 Semiconductor
UB1000
200828R4
P24
ZEISS
Sigma 300
목차
UB1000_P24 200828R4 Set up
타겟 장비
ZEISS - Sigma 300
설치 위치
N/A
비고
Good Performance
END USER
N/A
Operator: CHOI, HYOUNGMUN
Date of setup : 22.01.10
Date of reporting : 22.01.10
Date of business trip : 17.12.27~17.12.29
개요
After Active mounting, measure the vibration environment to check that there is no problem in using the equipment.
Measurement date: 22.01.10(Mon)
Operator: Daeil System, Research engineer CHOI, HYOUNGMUN
Place of measurement :
N/A
Equipment Model: ZEISS-Sigma 300
측정 조건 및 결과
측정 결과 요약
| 축 | Transmissibility (5 Hz) | Transmissibility (10 Hz) |
|---|---|---|
| Z | -17 dB | -25 dB |
| X | -16 dB | -25 dB |
| Y | -17 dB | -25 dB |
전방향 성능 양호
Good omnidirectional performance.
UB1000 내부 프로그램 성능 측정 결과
Z축
X축
Y축
Reference
일반 진동 기준
참고:
- VC-A/B는 8-80 Hz, VC-C~VC-G는 1-80 Hz 범위의 1/3 옥타브 밴드에서 측정.
- 세부 크기는 마이크로일렉트로닉스 제조에서의 선폭 또는 의료 연구에서의 입자 크기를 의미.