설치 보고서
DVIA-P Series
05-26-2026
Guangzhou Bio Island Thermoscientific Axia UB700 P3 (181107R4-2) Installation Report
DVIA-P Series
Installation Report
Guangzhou Bio Island
Thermoscientific
Axia
UB700
181107R4-2
Transmissibility
internal program
목차
워크시트
UB700 P3(181107R4-2)
대상 장비
Thermoscientific Axia
설치 장소
guangzhou
비고
Good Performance
End User
广州生物岛
운전자 · 설치 · 보고
Operator: Kim, Doohyun
Date of setup
21.10.25
Date of reporting
21.10.25
Date of business trip
17.12.27~17.12.29
개요
After Active mounting, measure the vibration environment to check that there is no problem in using the equipment.
측정 일자
Measurement date: 21.10.25 (MON)
측정 담당
Operator: Daeil System, Research engineer Kim, Doohyun
측정 장소
Place of measurement : guangzhou, 广州生物岛
Equipment Model
Thermoscientific Axia
측정 조건 및 결과
측정 결과 요약
| Test Condition | 2 Hz | 3 Hz | 10 Hz |
|---|---|---|---|
| Z | -3.5 dB | -10 dB | -28.2 dB |
| X | -3.5 dB | -10 dB | -25.8 dB |
| Y | -3.7 dB | -10 dB | -25.9 dB |
Good omnidirectional performance.
MB1000 internal program performance measurement result
Z축
X축
Y축