본문으로 바로가기
설치 보고서
DVIA-P Series
05-26-2026

Guangzhou Bio Island Thermoscientific Axia UB700 P3 (181107R4-2) Installation Report

DVIA-P Series
Installation Report
Guangzhou Bio Island
Thermoscientific
Axia
UB700
181107R4-2
Transmissibility
internal program

워크시트

UB700 P3(181107R4-2)

대상 장비

Thermoscientific Axia

설치 장소

guangzhou

비고

Good Performance

End User

广州生物岛

운전자 · 설치 · 보고

Operator: Kim, Doohyun

Date of setup

21.10.25

Date of reporting

21.10.25

Date of business trip

17.12.27~17.12.29

개요

After Active mounting, measure the vibration environment to check that there is no problem in using the equipment.

측정 일자

Measurement date: 21.10.25 (MON)

측정 담당

Operator: Daeil System, Research engineer Kim, Doohyun

측정 장소

Place of measurement : guangzhou, 广州生物岛

Equipment Model

Thermoscientific Axia

측정 조건 및 결과

측정 결과 요약

Test Condition2 Hz3 Hz10 Hz
Z-3.5 dB-10 dB-28.2 dB
X-3.5 dB-10 dB-25.8 dB
Y-3.7 dB-10 dB-25.9 dB

Good omnidirectional performance.

MB1000 internal program performance measurement result

Z축

X축

Y축

이 케이스 스터디 공유하기

케이스 스터디 정보

카테고리
설치 보고서
시리즈DVIA-P Series
작성일05-26-2026
태그
DVIA-P Series
Installation Report
Guangzhou Bio Island
Thermoscientific
Axia
UB700
181107R4-2
Transmissibility
internal program