Wenzhou Medical College JEOL F200 TEM DVIA-M3000 re-tuning report
DVIA-ML Series
Installation Report
JEOL
TEM
F200
목차
Wenzhou Medical College
(M3000_ 200916R1 Re-Tuning)
| 대상 장비 | JEOL F200 TEM |
|---|---|
| 설치 장소 | WENZHOU |
| 비고 | Occurrence of x-axis vibration amplification |
| END USER | Wenzhou Medical College |
| Operator: Young ha, Lee Date of setup: 22.06.09 Date of reporting: 22.06.09 | Date of business trip : 17.12.27~17.12.29 |
개요
The DVIA-M3000 active vibration isolation platform has been retuned, but it has not yielded as much performance as desired. Intermittent amplification of the 11Hz band occurs on the X-axis. It is difficult to distinguish whether the vibrations in the 11Hz band are occurring in an active vibration isolator or on the equipment at the top, due to the limitations of remote tuning. Needs to measure if the top plate experiences vibrations in the 11 Hz band in the Active Off state. Measurement date: 2022.06.09 (Thursday), Operator: DAEIL SYSTEMS, Youngha Lee Place of measurement: WENZHOU Equipment Model: JEOL F200 TEM
M3000 internal program performance measurement result
Z axis
X axis
Amplification occurs in the 11 Hz band on the X axis
Y axis
이 케이스 스터디 공유하기
케이스 스터디 정보
카테고리
설치 보고서
시리즈DVIA-ML Series
작성일06-09-2022
태그
DVIA-ML Series
Installation Report
JEOL
TEM
F200
관련 케이스 스터디

DVIA-ML Series
Ansteel Iron and Steel Research Institute Thermo Scientific Talos TEM DVIA-ML3000 (251125R4) Installation Report
07-23-2026자세히

DVIA-ML Series
Guangzhou GRG Thermo Fisher Talos TEM DVIA-ML3000 (260128R4) Installation Report
04-22-2026자세히

DVIA-ML Series
Institute of Geodesy and Geophysics Thermo Fisher TEM Spectra 300 DVIA-ML3000 Installation Report
06-19-2025자세히